![]() ![]() We recommend placing a Biological Indicator in these locations to confirm kill. This is caused by paper absorbing the moisture from the steam. ![]() Localised Dry Heat – High temperatures are often seen by thermocouples placed on paper and mop heads.Superheated Steam – Perform Steam Quality Testing Superheat test.Calibration – Has this been performed recently and the correct off sets applied?.If excursions are identified then there are several potential route causes: The theoretical temperature can be correlated by the pressure recorded by using steam table for example Ĭontrol of temperatures and pressures within the autoclave should fall within the control set points and correlate to steam tables. Assuming that all air has been removed from the autoclave chamber prior to the sterilisation phase the temperature and pressure should correlate. Porous load autoclaves usually control by either temperature or pressure.